Five-year follow-up data from the U.S. clinical trial for Sientra's U.S. Food and Drug Administration-approved Silimed® brand round and shaped implants with high-strength silicone gel.
Abstract
[BACKGROUND] In March of 2012, the U.S. Food and Drug Administration approved Sientra's application for premarket approval for its Silimed brand silicone gel implants, based on data from the largest silicone gel breast implant study to date. This was the first approval for shaped silicone gel breast implants. This article presents the results of Sientra's study through 5 years.
[METHODS] Sientra's study is an ongoing, 10-year, open-label, prospective, multicenter clinical study designed to assess the safety and effectiveness of Sientra's implants in patients undergoing augmentation and reconstruction. A total of 1788 subjects were implanted with 3506 implants, including 1116 primary augmentation, 363 revision-augmentation, 225 primary reconstruction, and 84 revision-reconstruction subjects. Physical evaluations and complications were recorded at each visit. Effectiveness was measured by postimplantation bra cup size and assessment of subject satisfaction and quality of life. Of the 1788 subjects, 571 underwent magnetic resonance imaging to assess silent rupture. Safety endpoints were analyzed using the Kaplan-Meier method.
[RESULTS] Across all cohorts, the risk of rupture was 1.8 percent (95 percent CI, 1.2 to 2.6 percent), the risk of capsular contracture (Baker grade III/IV) was 9.0 percent (95 percent CI, 7.6 to 10.6 percent), and the risk of reoperation was 23.8 percent (95 percent CI, 21.8 to 26.0 percent). Over 99 percent of surgeons reported satisfaction with the postoperative results, and subject satisfaction remained high 5 years after implantation.
[CONCLUSION] The 5-year results of Sientra's study continue to provide a comprehensive safety and effectiveness profile of Sientra's portfolio of Silimed brand shaped and round implants.
[CLINICAL QUESTION/LEVEL OF EVIDENCE] Therapeutic, IV.
[METHODS] Sientra's study is an ongoing, 10-year, open-label, prospective, multicenter clinical study designed to assess the safety and effectiveness of Sientra's implants in patients undergoing augmentation and reconstruction. A total of 1788 subjects were implanted with 3506 implants, including 1116 primary augmentation, 363 revision-augmentation, 225 primary reconstruction, and 84 revision-reconstruction subjects. Physical evaluations and complications were recorded at each visit. Effectiveness was measured by postimplantation bra cup size and assessment of subject satisfaction and quality of life. Of the 1788 subjects, 571 underwent magnetic resonance imaging to assess silent rupture. Safety endpoints were analyzed using the Kaplan-Meier method.
[RESULTS] Across all cohorts, the risk of rupture was 1.8 percent (95 percent CI, 1.2 to 2.6 percent), the risk of capsular contracture (Baker grade III/IV) was 9.0 percent (95 percent CI, 7.6 to 10.6 percent), and the risk of reoperation was 23.8 percent (95 percent CI, 21.8 to 26.0 percent). Over 99 percent of surgeons reported satisfaction with the postoperative results, and subject satisfaction remained high 5 years after implantation.
[CONCLUSION] The 5-year results of Sientra's study continue to provide a comprehensive safety and effectiveness profile of Sientra's portfolio of Silimed brand shaped and round implants.
[CLINICAL QUESTION/LEVEL OF EVIDENCE] Therapeutic, IV.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 해부 | breast
|
유방 | dict | 2 | |
| 합병증 | breast implant
|
scispacy | 1 | ||
| 합병증 | capsular contracture
|
피막구축 | dict | 1 | |
| 약물 | high-strength silicone
|
scispacy | 1 | ||
| 약물 | silicone
|
C0037114
silicones
|
scispacy | 1 | |
| 약물 | [BACKGROUND]
|
scispacy | 1 | ||
| 질환 | rupture
|
C3203359
Rupture
|
scispacy | 1 | |
| 기타 | Silimed
|
scispacy | 1 | ||
| 기타 | patients
|
scispacy | 1 | ||
| 기타 | subjects
|
scispacy | 1 |
MeSH Terms
Breast Implantation; Breast Implants; Device Approval; Follow-Up Studies; Humans; Implant Capsular Contracture; Patient Satisfaction; Prosthesis Design; Prosthesis Failure; Quality of Life; Reoperation; Rupture
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- The impact of three-dimensional simulation and virtual reality technologies on surgical decision-making and postoperative satisfaction in aesthetic surgery: a preliminary study.
- Cutaneous fistula of the breast: A complication of cosmetic autologous fat transfer.
- Epidermal inclusion cyst after breast reduction mammoplasty.
- The Plastic Surgery In-Service Examination: A Scoping Review.
- Clinical outcomes of synthetic absorbable mesh use in breast surgery: First case series in reconstruction and aesthetic mastopexy.