Biomechanical comparison of titanium miniplates versus a variety of CAD/CAM plates in mandibular reconstruction.
Abstract
[BACKGROUND] Titanium plate fixation of free flaps in mandibular reconstruction involves complications such as osseous non-union or imaging artefacts. Interosteotomy movement (IOM) is known to affect bone healing. This study aimed to compare IOM and mechanical integrity of four different fixation systems in a mandible reconstruction model.
[METHODS] Two polyurethane (PU) fibula segments were fixed in right-sided defects of PU mandibles. Laser-melted patient-specific titanium plates were fixed with non-locking-screws (Ti-NL) or locking-screws (Ti-L). The third group consisted of locking-screws for patient-specific polyetheretherketone (PEEK-L) plates. The last group used titanium miniplates and monocortical screw fixation (Ti-MP). All models were loaded unilaterally via cyclic dynamic loading with increasing loads to simulate mastication. IOM was registered using a 3D optical tracking system.
[FINDINGS] PEEK-L showed highest vertical displacement (p = 0.010), lowest stiffness (p = 0.004) and highest IOM (p = 0.001). All specimen in PEEK-L demonstrated abnormal bending (n = 5) or plate fracture (n = 1). Vertical displacement or stiffness did not differ between any of Ti-MP, Ti-L and Ti-NL. IOM in Ti-MP was higher than in Ti-L and Ti-NL (p = 0.001).
[INTERPRETATION] Mechanical integrity of all titanium plates complies with established standards. In this model, the screw system did not influence IOM. In the tested composition and shape, PEEK plates do not seem to guarantee sufficient mechanical integrity for a use in mandibular reconstruction. Thus modifications are needed. Future clinical studies are needed to clarify optimal IOM after mandible reconstruction.
[METHODS] Two polyurethane (PU) fibula segments were fixed in right-sided defects of PU mandibles. Laser-melted patient-specific titanium plates were fixed with non-locking-screws (Ti-NL) or locking-screws (Ti-L). The third group consisted of locking-screws for patient-specific polyetheretherketone (PEEK-L) plates. The last group used titanium miniplates and monocortical screw fixation (Ti-MP). All models were loaded unilaterally via cyclic dynamic loading with increasing loads to simulate mastication. IOM was registered using a 3D optical tracking system.
[FINDINGS] PEEK-L showed highest vertical displacement (p = 0.010), lowest stiffness (p = 0.004) and highest IOM (p = 0.001). All specimen in PEEK-L demonstrated abnormal bending (n = 5) or plate fracture (n = 1). Vertical displacement or stiffness did not differ between any of Ti-MP, Ti-L and Ti-NL. IOM in Ti-MP was higher than in Ti-L and Ti-NL (p = 0.001).
[INTERPRETATION] Mechanical integrity of all titanium plates complies with established standards. In this model, the screw system did not influence IOM. In the tested composition and shape, PEEK plates do not seem to guarantee sufficient mechanical integrity for a use in mandibular reconstruction. Thus modifications are needed. Future clinical studies are needed to clarify optimal IOM after mandible reconstruction.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 해부 | mandible
|
하악골 | dict | 2 | |
| 해부 | mandibular
|
scispacy | 1 | ||
| 해부 | flaps
|
scispacy | 1 | ||
| 해부 | bone
|
scispacy | 1 | ||
| 합병증 | osseous
|
scispacy | 1 | ||
| 약물 | Ti-MP
→ titanium miniplates and monocortical screw fixation
|
scispacy | 1 | ||
| 약물 | PEEK
|
C0084113
polyetheretherketone
|
scispacy | 1 | |
| 약물 | [BACKGROUND] Titanium
|
scispacy | 1 | ||
| 약물 | titanium
|
C0040302
titanium
|
scispacy | 1 | |
| 약물 | polyetheretherketone
|
C0084113
polyetheretherketone
|
scispacy | 1 | |
| 질환 | Ti-MP
→ titanium miniplates and monocortical screw fixation
|
scispacy | 1 | ||
| 질환 | abnormal bending
|
scispacy | 1 | ||
| 질환 | fracture
|
C0016658
Fracture
|
scispacy | 1 | |
| 질환 | osseous non-union
|
scispacy | 1 | ||
| 기타 | mandibular
|
scispacy | 1 | ||
| 기타 | PU mandibles
|
scispacy | 1 | ||
| 기타 | Ti-MP
→ titanium miniplates and monocortical screw fixation
|
scispacy | 1 |
MeSH Terms
Biomechanical Phenomena; Bone Plates; Fracture Fixation, Internal; Humans; Mandibular Reconstruction; Titanium
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Evaluation of the ultrasonography-guidance for botulinum toxin type a injection into lateral pterygoid muscle in patients with TMJ anterior disc displacement with reduction.
- What Lies Beneath: Carotid Artery Depth and the Safe Use of Hemostatic Nets in Facelift Surgery.
- Bony Union in Maxilla and Mandible Free Flap Reconstruction: A Systematic Review and Meta-Analysis.
- Long-term outcomes of CAD/CAM monocompartmental condyle prosthesis supporting bone transplant after mandible resection.
- From Bench to Bedside: A Pilot Clinical Study on NANOTEX Bone Graft for Fibula Onlay Augmentation in Mandibular Reconstruction.