Venous Flow Coupler in Head and Neck Free Flap Reconstruction.
Abstract
[OBJECTIVE] To describe the utility of venous flow couplers in monitoring free tissue flaps in the immediate postoperative setting.
[STUDY DESIGN] Retrospective case series.
[SETTING] Otolaryngology department at a single tertiary care institution.
[METHODS] A retrospective case series of free flap reconstructions in which venous flow couplers were employed to supplement flap monitoring. All free flap cases performed over the past 4 years were reviewed. Inclusion criteria were venous flow coupler and arterial flow Doppler monitored for 5 days postoperatively.
[RESULTS] From July 2014 through May 2018, the venous flow coupler was used with the arterial flow Doppler and clinical monitoring in 228 cases. Eleven cases did not meet criteria for inclusion; thus, 217 cases were analyzed. Twenty cases (9.2%) returned to the operating room with concern for flap compromise, and 16 were salvaged. The combination of venous flow coupler and arterial flow Doppler identified 19 of these flaps. Venous flow couplers identified 5 compromised flaps before there was an arterial signal change, and all were salvaged. Additionally, there was a 24.1% false-positive rate when 2 venous flow couplers were used in parallel. For the venous flow coupler, the positive predictive value was 64.3% and the negative predictive value, 98.9%. The false-positive rate in the series was 5.1%. The sensitivity was 90% and the specificity, 94.9%.
[CONCLUSION] The venous flow coupler is able to detect venous thrombosis in the absence of arterial thrombosis and may contribute to improved flap salvage rates.
[STUDY DESIGN] Retrospective case series.
[SETTING] Otolaryngology department at a single tertiary care institution.
[METHODS] A retrospective case series of free flap reconstructions in which venous flow couplers were employed to supplement flap monitoring. All free flap cases performed over the past 4 years were reviewed. Inclusion criteria were venous flow coupler and arterial flow Doppler monitored for 5 days postoperatively.
[RESULTS] From July 2014 through May 2018, the venous flow coupler was used with the arterial flow Doppler and clinical monitoring in 228 cases. Eleven cases did not meet criteria for inclusion; thus, 217 cases were analyzed. Twenty cases (9.2%) returned to the operating room with concern for flap compromise, and 16 were salvaged. The combination of venous flow coupler and arterial flow Doppler identified 19 of these flaps. Venous flow couplers identified 5 compromised flaps before there was an arterial signal change, and all were salvaged. Additionally, there was a 24.1% false-positive rate when 2 venous flow couplers were used in parallel. For the venous flow coupler, the positive predictive value was 64.3% and the negative predictive value, 98.9%. The false-positive rate in the series was 5.1%. The sensitivity was 90% and the specificity, 94.9%.
[CONCLUSION] The venous flow coupler is able to detect venous thrombosis in the absence of arterial thrombosis and may contribute to improved flap salvage rates.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | free flap
|
피판재건술 | dict | 3 | |
| 시술 | flap
|
피판재건술 | dict | 3 | |
| 합병증 | flaps
|
scispacy | 1 | ||
| 약물 | [OBJECTIVE]
|
scispacy | 1 | ||
| 질환 | venous thrombosis
|
C0042487
Venous Thrombosis
|
scispacy | 1 | |
| 질환 | thrombosis
|
C0040053
Thrombosis
|
scispacy | 1 | |
| 질환 | Head and Neck Free Flap
|
scispacy | 1 | ||
| 기타 | tissue flaps
|
scispacy | 1 | ||
| 기타 | arterial
|
scispacy | 1 | ||
| 기타 | venous
|
scispacy | 1 |
MeSH Terms
Aged; Female; Free Tissue Flaps; Head and Neck Neoplasms; Humans; Male; Middle Aged; Monitoring, Physiologic; Postoperative Care; Plastic Surgery Procedures; Regional Blood Flow; Retrospective Studies; Ultrasonography, Doppler
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Endodontic implications of hypercementosis: A systematic review of anatomical challenges and therapeutic strategies.
- Breast plastic surgery in perimenopausal and postmenopausal women: Menopause-informed counseling on screening, safety, and long-term breast health.
- Application of the SCIA-Pure Skin Perforator Flap in Bilateral Upper Eyelid Reconstruction: A Case Report and Review of the Literature.
- Free flap reconstruction of a cast-related pressure ulcer in a pediatric patient with spinal muscular atrophy.
- Characterization of Trimmed Nerve Morphology Using High-Resolution Imaging: Comparison of Three Surgical Instruments.