[OBJECTIVES] The aim of this study was to investigate the selected biomechanical properties of semi-anatomic implant plate made of biostable glass fiber-reinforced composite (GFRC) for mandibular reconstruction. Two versions of GFRC plates were tested in vitro loading conditions of a mandible segmental defect model, for determining the level of mechanical stress at the location of fixation screws, and in the body of the plate.

[METHODS] GFRC of bidirectional S3-glass fiber weaves with dimethacrylate resin matrix were used to fabricate semi-anatomic reconstruction plates of two GFRC laminate thicknesses. Lateral surface of the plate followed the contour of the resected part of the bone, and the medial surface was concave allowing for placement of a microvascular bone flap in the next stages of the research. Plates were fixed with screws to a plastic model of the mandible with a large segmental defect in the premolar-molar region. The mandible-plate system was loaded from incisal and molar locations with loads of 10, 50, and 100 N and stress (microstrain, με) at the location of fixation screws and the body of the plate was measured by strain gauges. In total the test set-up had four areas for measuring the stress of the plate.

[RESULTS] No signs of fractures or buckling failures of the plates were found during loading. Strain values at the region of the fixation screws were higher with thick plate, whereas thin plates demonstrated higher strain at the body of the plate. Vertical displacement of the mandible-plate system was proportional to the loading force and was higher with incisal than molar loading locations but no difference was found between thin and thick plates.

[CONCLUSION] GFRC plates withstood the loading conditions up to 100 N even when loaded incisally. Thick plates concentrated the stress to the ramus mandibulae region of the fixation screws whereas the thin plates showed stress concentration in the angulus mandibulae region of the fixation and the plate itself. In general, thin plates caused a lower magnitude of stress to the fixation screw areas than thick plates, suggesting absorption of the loading energy to the body of the plate.

추출된 의학 개체 (NER)

유형영어 표현한국어 / 풀이UMLS CUI출처등장
해부 mandible 하악골 dict 4
시술 microvascular 미세수술 dict 1
시술 flap 피판재건술 dict 1
해부 S3-glass fiber scispacy 1
해부 bone scispacy 1
해부 medial surface scispacy 1
해부 microvascular bone scispacy 1
해부 molar scispacy 1
해부 angulus mandibulae scispacy 1
약물 GFRC → glass fiber-reinforced scispacy 1
약물 dimethacrylate resin C0440141
Dimethacrylate resin
scispacy 1
약물 GFRC laminate scispacy 1
약물 N and stress scispacy 1
약물 [OBJECTIVES] scispacy 1
약물 dimethacrylate resin matrix scispacy 1
질환 mandibular segmental defects scispacy 1
질환 fractures C0016658
Fracture
scispacy 1
질환 GFRC → glass fiber-reinforced scispacy 1
기타 mandibular scispacy 1
기타 GFRC laminate scispacy 1
기타 premolar-molar scispacy 1
기타 incisal scispacy 1
기타 mandible-plate scispacy 1
기타 ramus mandibulae scispacy 1

MeSH Terms

Glass; Mandible; Biomechanical Phenomena; Materials Testing; Stress, Mechanical; Bone Plates; Mechanical Tests; Mechanical Phenomena; Humans

🔗 함께 등장하는 도메인

이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들

관련 논문