A Comparative Evaluation of Three Different Attachment Methods for a Video Laryngoscope in a Surgical Setting.
Abstract
[INTRODUCTION] The prerequisite for transoral microsurgery of laryngeal lesions is an uninterrupted line of sight to the operative field. Patients with cervical stiffness or anatomical variations that prevent adequate laryngeal exposure are typically unsuitable for this procedure. In such cases, a curved video laryngoscope may facilitate improved access to the larynx. However, conventional suspension laryngoscopy setups are incompatible with standard video laryngoscopes.
[OBJECTIVE] This study evaluates three attachment methods for integrating a curved video laryngoscope into a surgical setup, focusing on their resistance to external forces.
[METHODS] This study assessed three different attachment methods (3D-printed clamp vs. metal bracket vs. articulated stand) for a video laryngoscope in a surgical setup. External forces, both lateral and rotational (torque), were applied and continuously measured until laryngeal visualization was compromised by displacement of the video laryngoscope.
[RESULTS] The metal bracket demonstrated significantly ( < 0.001) higher resistance to lateral forces (median 184.49 N, 95% CI [181.59-189.61 N]) compared to the articulated stand (median 88.16 N, 95% CI [76.73-88.98 N]) and the 3D-printed clamp (median 55.59 N, 95% CI [54.74-57.58 N]). The articulated stand exhibited significantly ( < 0.005) greater torque resistance (median 9.57 N m, 95% CI [5.65-9.87 N m]) compared to the metal bracket (median 1.58 N m, 95% CI [1.57-2.13 N m]) and the 3D-printed clamp (median 2.46 N m, 95% CI [2.24-2.79 N m]).
[CONCLUSION] Overall, the articulated stand outperformed the other attachment methods, displaying robust resistance to lateral forces and superior rotational stability.
[LEVEL OF EVIDENCE] Level 4.
[OBJECTIVE] This study evaluates three attachment methods for integrating a curved video laryngoscope into a surgical setup, focusing on their resistance to external forces.
[METHODS] This study assessed three different attachment methods (3D-printed clamp vs. metal bracket vs. articulated stand) for a video laryngoscope in a surgical setup. External forces, both lateral and rotational (torque), were applied and continuously measured until laryngeal visualization was compromised by displacement of the video laryngoscope.
[RESULTS] The metal bracket demonstrated significantly ( < 0.001) higher resistance to lateral forces (median 184.49 N, 95% CI [181.59-189.61 N]) compared to the articulated stand (median 88.16 N, 95% CI [76.73-88.98 N]) and the 3D-printed clamp (median 55.59 N, 95% CI [54.74-57.58 N]). The articulated stand exhibited significantly ( < 0.005) greater torque resistance (median 9.57 N m, 95% CI [5.65-9.87 N m]) compared to the metal bracket (median 1.58 N m, 95% CI [1.57-2.13 N m]) and the 3D-printed clamp (median 2.46 N m, 95% CI [2.24-2.79 N m]).
[CONCLUSION] Overall, the articulated stand outperformed the other attachment methods, displaying robust resistance to lateral forces and superior rotational stability.
[LEVEL OF EVIDENCE] Level 4.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | microsurgery
|
미세수술 | dict | 1 | |
| 해부 | cervical
|
scispacy | 1 | ||
| 해부 | laryngeal
|
scispacy | 1 | ||
| 해부 | larynx
|
scispacy | 1 | ||
| 해부 | lateral
|
scispacy | 1 | ||
| 해부 | CI [181.59-189.61
|
scispacy | 1 | ||
| 합병증 | laryngeal lesions
|
scispacy | 1 | ||
| 약물 | CI [1.57-
|
scispacy | 1 | ||
| 약물 | CI [
|
scispacy | 1 | ||
| 약물 | [INTRODUCTION] The
|
scispacy | 1 | ||
| 약물 | [OBJECTIVE]
|
scispacy | 1 | ||
| 약물 | CI [76.73-88.98
|
scispacy | 1 | ||
| 약물 | CI [54.74-57.58
|
scispacy | 1 | ||
| 질환 | cervical stiffness
|
scispacy | 1 | ||
| 기타 | Patients
|
scispacy | 1 |
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Endodontic implications of hypercementosis: A systematic review of anatomical challenges and therapeutic strategies.
- Breast plastic surgery in perimenopausal and postmenopausal women: Menopause-informed counseling on screening, safety, and long-term breast health.
- Application of the SCIA-Pure Skin Perforator Flap in Bilateral Upper Eyelid Reconstruction: A Case Report and Review of the Literature.
- Free flap reconstruction of a cast-related pressure ulcer in a pediatric patient with spinal muscular atrophy.
- Characterization of Trimmed Nerve Morphology Using High-Resolution Imaging: Comparison of Three Surgical Instruments.