Design of a novel pH sensor for flap monitoring and evaluation of its efficacy in an in vivo flap model.
Abstract
[PURPOSE] This study aimed to design a novel pH sensor for free flap monitoring and to evaluate its efficacy in a preliminary animal model, with the goal of supporting earlier detection of perfusion changes.
[MATERIALS AND METHODS] The study was conducted in two phases. First, the sensor was designed, fabricated, and validated through in vitro testing. In the second phase, 15 Sprague-Dawley rats were divided into three groups based on the type of vascular occlusion: arterial, venous, or total pedicle. A superficial inferior epigastric island flap was elevated in each animal, and the sensor was placed beneath the flap. Continuous measurements were recorded for baseline, occlusion, recovery, and repeat occlusion phases, each lasting 20 minutes.
[RESULTS] Arterial and venous occlusions produced characteristic current signal peaks, whereas pedicle occlusion resulted in irregular, noisy readings. The mean response times were 104.8, 130, and 170 seconds for the arterial, venous, and pedicle groups, respectively.
[CONCLUSION] In this pilot study, the novel pH sensor detected changes in current output following vascular occlusion in a rat flap model. These findings suggest potential for use in free flap monitoring. Further development is needed to produce biocompatible, wireless, and clinically applicable versions for human use.
[MATERIALS AND METHODS] The study was conducted in two phases. First, the sensor was designed, fabricated, and validated through in vitro testing. In the second phase, 15 Sprague-Dawley rats were divided into three groups based on the type of vascular occlusion: arterial, venous, or total pedicle. A superficial inferior epigastric island flap was elevated in each animal, and the sensor was placed beneath the flap. Continuous measurements were recorded for baseline, occlusion, recovery, and repeat occlusion phases, each lasting 20 minutes.
[RESULTS] Arterial and venous occlusions produced characteristic current signal peaks, whereas pedicle occlusion resulted in irregular, noisy readings. The mean response times were 104.8, 130, and 170 seconds for the arterial, venous, and pedicle groups, respectively.
[CONCLUSION] In this pilot study, the novel pH sensor detected changes in current output following vascular occlusion in a rat flap model. These findings suggest potential for use in free flap monitoring. Further development is needed to produce biocompatible, wireless, and clinically applicable versions for human use.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | flap
|
피판재건술 | dict | 5 | |
| 시술 | free flap
|
피판재건술 | dict | 2 | |
| 합병증 | vascular occlusion
|
혈관폐색 | dict | 2 | |
| 해부 | pedicle
|
scispacy | 1 | ||
| 질환 | pedicle occlusion
|
scispacy | 1 | ||
| 기타 | Sprague-Dawley rats
|
scispacy | 1 | ||
| 기타 | vascular
|
scispacy | 1 | ||
| 기타 | arterial
|
scispacy | 1 | ||
| 기타 | venous
|
scispacy | 1 | ||
| 기타 | superficial inferior epigastric
|
scispacy | 1 | ||
| 기타 | rat flap
|
scispacy | 1 | ||
| 기타 | human
|
scispacy | 1 |
MeSH Terms
Animals; Rats, Sprague-Dawley; Rats; Free Tissue Flaps; Hydrogen-Ion Concentration; Monitoring, Physiologic; Models, Animal; Equipment Design; Pilot Projects; Male
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Endodontic implications of hypercementosis: A systematic review of anatomical challenges and therapeutic strategies.
- Breast plastic surgery in perimenopausal and postmenopausal women: Menopause-informed counseling on screening, safety, and long-term breast health.
- Application of the SCIA-Pure Skin Perforator Flap in Bilateral Upper Eyelid Reconstruction: A Case Report and Review of the Literature.
- Free flap reconstruction of a cast-related pressure ulcer in a pediatric patient with spinal muscular atrophy.
- Characterization of Trimmed Nerve Morphology Using High-Resolution Imaging: Comparison of Three Surgical Instruments.