Microfocused Ultrasound With Visualization: A Systematic Review of Adverse Events and Risk of Subsequent Facelift Compromise.
Abstract
[BACKGROUND] Plastic surgery colleagues have expressed unconfirmed concerns regarding the deleterious implications of noninvasive energy-based therapies for skin laxity, such as microfocused ultrasound with visualization (MFU-V), on preexisting facial fat or subsequent surgical facelifts.
[OBJECTIVE] The authors conducted a systematic review to summarize treatment-related adverse events (TRAEs) following MFU-V and elucidate their impact on future rhytidectomies.
[MATERIALS AND METHODS] The authors performed a PubMed/MEDLINE search using the keywords "MFU-V," "fat loss," "adverse events," "facelift," and their derivatives. In addition, the authors searched the manufacturer and user facility device experience (MAUDE) database using the brand name "Ulthera."
[RESULTS] Nineteen articles with data on 506 patients were identified. Transient edema, erythema, and postprocedural pain were observed most often. Instances of significant TRAEs were limited in the scientific literature, apart from 1 case of subcutaneous atrophy. Data from 106 records were extracted from the MAUDE database. Notably, lipoatrophy, neurologic sequelae (including nerve damage, focal numbness, dysesthesia, and ptosis), and scarring were reported with the most frequency. There was 1 report of facelift compromise.
[CONCLUSION] While the scientific literature lacks supporting evidence regarding the impact of MFU-V on subsequent facelift procedures, anecdotal, largely unverified reports exist. Hence, peer-reviewed literature detailing such occurrences would be valuable in understanding the potential ensuing risks.
[OBJECTIVE] The authors conducted a systematic review to summarize treatment-related adverse events (TRAEs) following MFU-V and elucidate their impact on future rhytidectomies.
[MATERIALS AND METHODS] The authors performed a PubMed/MEDLINE search using the keywords "MFU-V," "fat loss," "adverse events," "facelift," and their derivatives. In addition, the authors searched the manufacturer and user facility device experience (MAUDE) database using the brand name "Ulthera."
[RESULTS] Nineteen articles with data on 506 patients were identified. Transient edema, erythema, and postprocedural pain were observed most often. Instances of significant TRAEs were limited in the scientific literature, apart from 1 case of subcutaneous atrophy. Data from 106 records were extracted from the MAUDE database. Notably, lipoatrophy, neurologic sequelae (including nerve damage, focal numbness, dysesthesia, and ptosis), and scarring were reported with the most frequency. There was 1 report of facelift compromise.
[CONCLUSION] While the scientific literature lacks supporting evidence regarding the impact of MFU-V on subsequent facelift procedures, anecdotal, largely unverified reports exist. Hence, peer-reviewed literature detailing such occurrences would be valuable in understanding the potential ensuing risks.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | facelift
|
안면거상술 | dict | 4 | |
| 해부 | subcutaneous
|
피하조직 | dict | 1 | |
| 해부 | skin
|
scispacy | 1 | ||
| 해부 | facial fat
|
scispacy | 1 | ||
| 해부 | fat
|
scispacy | 1 | ||
| 해부 | nerve
|
scispacy | 1 | ||
| 해부 | focal
|
scispacy | 1 | ||
| 합병증 | edema
|
scispacy | 1 | ||
| 합병증 | erythema
|
scispacy | 1 | ||
| 약물 | 506
|
scispacy | 1 | ||
| 약물 | [BACKGROUND] Plastic surgery
|
scispacy | 1 | ||
| 약물 | [OBJECTIVE]
|
scispacy | 1 | ||
| 질환 | TRAEs
→ treatment-related adverse events
|
scispacy | 1 | ||
| 질환 | fat loss
|
scispacy | 1 | ||
| 질환 | edema
|
C0013604
Edema
|
scispacy | 1 | |
| 질환 | erythema
|
C0041834
Erythema
|
scispacy | 1 | |
| 질환 | pain
|
C0030193
Pain
|
scispacy | 1 | |
| 질환 | atrophy
|
C0333641
Atrophic
|
scispacy | 1 | |
| 질환 | neurologic sequelae
|
C0746864
Neurologic sequelae
|
scispacy | 1 | |
| 질환 | nerve damage
|
C0161479
Nerve injury
|
scispacy | 1 | |
| 질환 | numbness
|
C0020580
Hypesthesia
|
scispacy | 1 | |
| 질환 | dysesthesia
|
C0392699
Dysesthesia
|
scispacy | 1 | |
| 질환 | MFU-V
→ microfocused ultrasound with visualization
|
scispacy | 1 | ||
| 기타 | MFU-V
→ microfocused ultrasound with visualization
|
scispacy | 1 | ||
| 기타 | patients
|
scispacy | 1 |
MeSH Terms
Humans; Rhytidoplasty; Skin Aging; Ultrasonic Therapy; Postoperative Complications
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Implications of Dermatologic Disorders in Facial Cosmetic Surgery: A Systematic Review.
- Clinical safety of a low-modification hyaluronic acid filler (MoD 2%) for facial rejuvenation.
- Implant-based versus autologous mastopexy after massive weight loss: Complications and patient satisfaction.
- Hyperhidrosis: Prevalence, Diagnosis, and Stepwise Treatment.
- Application of the SCIA-Pure Skin Perforator Flap in Bilateral Upper Eyelid Reconstruction: A Case Report and Review of the Literature.