Safety and tolerability of a focused ultrasound device for treatment of adipose tissue in subjects undergoing abdominoplasty: a placebo-control pilot study.
Abstract
[BACKGROUND] High-intensity focused ultrasound (HIFU) lipolysis is a noninvasive alternative to existing surgical body-sculpting methods.
[OBJECTIVE] To evaluate the safety, tolerability, and histologic outcome of HIFU lipolysis using a novel device in human subjects.
[METHODS AND MATERIALS] In a single-blind pilot study, six healthy subjects scheduled to undergo abdominoplasty within 4 weeks received HIFU lipolysis on one side of the umbilicus. An identical placebo treatment was given to the contralateral side. Patient evaluation of complications, blood tests, and urine analysis were performed 1, 3, 7, 14, and 28 days after treatment. Excised tissue from the treated areas was sent for histologic review.
[RESULTS] Treatment was well tolerated. Average visual analogue pain scale scores were 3.5 ± 2.3 (range 1-7) and 0.17 ± 0.41 (range 0-1). No major adverse events were documented, and laboratory analysis after HIFU lipolysis was normal. Fat necrosis with infiltration of lymphocytes and macrophages without adjacent tissue damage was documented on histology 2 to 4 weeks after HIFU lipolysis. Damage extent correlated with size of the area treated. No pathologic findings were found on the control side.
[CONCLUSIONS] High-intensity focused ultrasound treatment was well tolerated and safe. Focal damage to target tissue was documented, with adjacent tissues remaining intact.
[OBJECTIVE] To evaluate the safety, tolerability, and histologic outcome of HIFU lipolysis using a novel device in human subjects.
[METHODS AND MATERIALS] In a single-blind pilot study, six healthy subjects scheduled to undergo abdominoplasty within 4 weeks received HIFU lipolysis on one side of the umbilicus. An identical placebo treatment was given to the contralateral side. Patient evaluation of complications, blood tests, and urine analysis were performed 1, 3, 7, 14, and 28 days after treatment. Excised tissue from the treated areas was sent for histologic review.
[RESULTS] Treatment was well tolerated. Average visual analogue pain scale scores were 3.5 ± 2.3 (range 1-7) and 0.17 ± 0.41 (range 0-1). No major adverse events were documented, and laboratory analysis after HIFU lipolysis was normal. Fat necrosis with infiltration of lymphocytes and macrophages without adjacent tissue damage was documented on histology 2 to 4 weeks after HIFU lipolysis. Damage extent correlated with size of the area treated. No pathologic findings were found on the control side.
[CONCLUSIONS] High-intensity focused ultrasound treatment was well tolerated and safe. Focal damage to target tissue was documented, with adjacent tissues remaining intact.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | abdominoplasty
|
복부성형술 | dict | 2 | |
| 해부 | adipose tissue
|
scispacy | 1 | ||
| 해부 | blood
|
scispacy | 1 | ||
| 해부 | tissue
|
scispacy | 1 | ||
| 해부 | Fat
|
scispacy | 1 | ||
| 해부 | lymphocytes
|
scispacy | 1 | ||
| 해부 | macrophages
|
scispacy | 1 | ||
| 해부 | tissues
|
scispacy | 1 | ||
| 합병증 | necrosis
|
괴사 | dict | 1 | |
| 합병증 | umbilicus
|
scispacy | 1 | ||
| 약물 | [BACKGROUND] High-intensity
|
scispacy | 1 | ||
| 약물 | [OBJECTIVE]
|
scispacy | 1 | ||
| 약물 | urine
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS]
|
scispacy | 1 | ||
| 질환 | pain
|
C0030193
Pain
|
scispacy | 1 | |
| 질환 | tissue damage
|
C0010957
Tissue damage
|
scispacy | 1 | |
| 기타 | human
|
scispacy | 1 | ||
| 기타 | Patient
|
scispacy | 1 |
MeSH Terms
Abdominoplasty; Adipose Tissue; Adult; Equipment Design; Female; Humans; Lipectomy; Lipolysis; Male; Middle Aged; Pilot Projects; Single-Blind Method; Ultrasonography
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Case report of a rare soft tissue tuberculosis in a patient undergoing lipoabdominoplasty.
- What is the potential role of the nonopioid suzetrigine in pain management?
- Pedicled extensor digitorum brevis flap for the distal third lower limb reconstruction: A review of the literature.
- Ex Vivo and In Vivo Histological Evaluation of a 3-μm Wavelength, 40-μm Spot Size Fractional Laser System for Dermatology.
- "Immediate Free Nipple Grafting with Autologous Breast Reconstruction for Large and/or Ptotic Breasts: Preserving the "Nipple-Sparing" Option".