Central sterile services department: screening of automated cleaning in liposuction cannulae.
[OBJECTIVE] To evaluate the effectiveness of automated cleaning of 4 mm liposuction cannulas using ultrasonic washer machine, screening for dirt using a cannulate microscope device and the screening o
APA
Santos RKD, Pozzer CE, et al. (2022). Central sterile services department: screening of automated cleaning in liposuction cannulae.. Revista gaucha de enfermagem, 43, e20210057. https://doi.org/10.1590/1983-1447.2022.20210057.en
MLA
Santos RKD, et al.. "Central sterile services department: screening of automated cleaning in liposuction cannulae.." Revista gaucha de enfermagem, vol. 43, 2022, pp. e20210057.
PMID
36350960
Abstract
[OBJECTIVE] To evaluate the effectiveness of automated cleaning of 4 mm liposuction cannulas using ultrasonic washer machine, screening for dirt using a cannulate microscope device and the screening of viable strains of Staphylococcus aureus, with microbiological analysis.
[METHOD] Experimental study with 14 units of 4 mm liposuction cannulas performed at the Central Sterile Services Department and at the Microbiology Laboratory of a Hospital Complex, located in Porto Alegre, RS, Brazil, from October 2018 to April 2019.
[RESULTS] Of the 14 liposuction cannulas that constituted the sample, 42.9% remained dirty despite automated cleaning, and all of them showed recovery of viable microrganisms in the microbiological laboratory.
[CONCLUSION] The study showed that it is not possible to ensure the cleaning efficacy of automated cleaning of 4 mm liposuction cannulas, especially due to the conformation of the material with internal spaces with accumulation of dirt.
[METHOD] Experimental study with 14 units of 4 mm liposuction cannulas performed at the Central Sterile Services Department and at the Microbiology Laboratory of a Hospital Complex, located in Porto Alegre, RS, Brazil, from October 2018 to April 2019.
[RESULTS] Of the 14 liposuction cannulas that constituted the sample, 42.9% remained dirty despite automated cleaning, and all of them showed recovery of viable microrganisms in the microbiological laboratory.
[CONCLUSION] The study showed that it is not possible to ensure the cleaning efficacy of automated cleaning of 4 mm liposuction cannulas, especially due to the conformation of the material with internal spaces with accumulation of dirt.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | liposuction
|
지방흡입 | dict | 5 | |
| 합병증 | dirt
|
scispacy | 1 |
MeSH Terms
Humans; Lipectomy; Cannula; Brazil
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Quantitative Assessment of Cannula Kinematics in Liposuction Surgical Procedures Using a Marker-Based Tracking System.
- Fat harvesting protocol for enhanced stem cell viability - pilot study.
- Scarless Infragluteal Fixation (SIF): Correction of Post-Liposuction Infragluteal Deformity.
- [Super microsurgical lymphaticovenular anastomosis for limb lymphedema: An outcome analysis based on clinical stage and indocyanine green pattern].
- Iatrogenic pneumothorax associated with surgeries at anatomically thoracic-adjacent and non-adjacent sites: case report and scoping review.