Mean glandular dose in digital mamography in women with breast implants.
Abstract
[PURPOSE] Mean Glandular Dose (MGD) is the quantity related to the risk of radiation-induced breast cancer. This study aimed to evaluate the MGD in screening mammography for women with breast implants.
[METHODS] This retrospective study used data of 2680 mammographies of 335 asymptomatic women with mammary implants examined in a digital direct x-ray unit. Each woman had a total of eight images: four in standard views and four with posterior displacement of the implant (ID). Data on kV, mAs, target/filter combination, compressed breast thickness and MGD were obtained from the DICOM header of the stored images. Quantitative variables were presented through descriptive statistics for median (5th-95th percentiles); and the qualitative variables were presented by numbers and percentages. Mean glandular doses of standard views and ID views were compared and statistical analysis was used to assess the influence of implant position, breast glandularity and thickness on mean glandular doses.
[RESULTS] Median MGD for standard views were 3.30 (2.60-4.00) mGy for CC and 3.31 (2.70-4.20) mGy for MLO. For ID views, median were 1.20 (0.90-2.20) mGy and 1.40 (0.97-3.74) mGy for CC and MLO views, respectively. Median MGD for the whole examination of women with breast implants was 9.60 (7.92-12.07) mGy, ranging from 6.25 to 21.50 mGy. When comparing MGD median for standard and ID views it was found a statistically significant difference (p < 0.05), with higher doses for the standard views due to the greater compressed breast thickness in these views. It was observed that, in the standard views, MGD decreases with increasing breast thickness due to the manual radiographic techniques used to expose the women. It was also observed that implant position does not affect MGD in breast augmentation mammography.
[CONCLUSION] Mammography of women with mammary implants gives higher radiation doses when compared with those without implants. For more accurate dose assessment in augmented breast mammography, it is necessary that specific conversion factors for the calculation of MGD based on air kerma at entrance of breasts with implants are made available.
[METHODS] This retrospective study used data of 2680 mammographies of 335 asymptomatic women with mammary implants examined in a digital direct x-ray unit. Each woman had a total of eight images: four in standard views and four with posterior displacement of the implant (ID). Data on kV, mAs, target/filter combination, compressed breast thickness and MGD were obtained from the DICOM header of the stored images. Quantitative variables were presented through descriptive statistics for median (5th-95th percentiles); and the qualitative variables were presented by numbers and percentages. Mean glandular doses of standard views and ID views were compared and statistical analysis was used to assess the influence of implant position, breast glandularity and thickness on mean glandular doses.
[RESULTS] Median MGD for standard views were 3.30 (2.60-4.00) mGy for CC and 3.31 (2.70-4.20) mGy for MLO. For ID views, median were 1.20 (0.90-2.20) mGy and 1.40 (0.97-3.74) mGy for CC and MLO views, respectively. Median MGD for the whole examination of women with breast implants was 9.60 (7.92-12.07) mGy, ranging from 6.25 to 21.50 mGy. When comparing MGD median for standard and ID views it was found a statistically significant difference (p < 0.05), with higher doses for the standard views due to the greater compressed breast thickness in these views. It was observed that, in the standard views, MGD decreases with increasing breast thickness due to the manual radiographic techniques used to expose the women. It was also observed that implant position does not affect MGD in breast augmentation mammography.
[CONCLUSION] Mammography of women with mammary implants gives higher radiation doses when compared with those without implants. For more accurate dose assessment in augmented breast mammography, it is necessary that specific conversion factors for the calculation of MGD based on air kerma at entrance of breasts with implants are made available.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 해부 | breast
|
유방 | dict | 10 | |
| 해부 | mammary
|
유방 | dict | 2 | |
| 시술 | breast augmentation
|
유방성형술 | dict | 1 | |
| 해부 | glandular
|
scispacy | 1 | ||
| 해부 | posterior
|
scispacy | 1 | ||
| 해부 | MGD
→ Mean Glandular Dose
|
scispacy | 1 | ||
| 해부 | breasts
|
scispacy | 1 | ||
| 합병증 | MGD
→ Mean Glandular Dose
|
scispacy | 1 | ||
| 약물 | MGD
→ Mean Glandular Dose
|
scispacy | 1 | ||
| 약물 | thickness
|
C1280412
Thick
|
scispacy | 1 | |
| 약물 | [PURPOSE] Mean
|
scispacy | 1 | ||
| 질환 | breast cancer
|
C0006142
Malignant neoplasm of breast
|
scispacy | 1 | |
| 질환 | breast glandularity
|
scispacy | 1 | ||
| 질환 | breasts
|
C0006141
Breast
|
scispacy | 1 | |
| 질환 | MGD
→ Mean Glandular Dose
|
scispacy | 1 | ||
| 질환 | MLO
|
scispacy | 1 | ||
| 기타 | women
|
scispacy | 1 | ||
| 기타 | Glandular
|
scispacy | 1 | ||
| 기타 | MLO
|
scispacy | 1 |
MeSH Terms
Adult; Breast Implants; Female; Humans; Mammary Glands, Human; Mammography; Middle Aged; Radiation Dosage; Retrospective Studies
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- The impact of three-dimensional simulation and virtual reality technologies on surgical decision-making and postoperative satisfaction in aesthetic surgery: a preliminary study.
- Cutaneous fistula of the breast: A complication of cosmetic autologous fat transfer.
- Epidermal inclusion cyst after breast reduction mammoplasty.
- Clinical outcomes of synthetic absorbable mesh use in breast surgery: First case series in reconstruction and aesthetic mastopexy.
- Implant-based versus autologous mastopexy after massive weight loss: Complications and patient satisfaction.