Clinical Results of Mentor MemoryGel Xtra Breast Implants From the GLOW Clinical Trial.
Abstract
[BACKGROUND] Mentor MemoryGel Xtra breast implants (Mentor Worldwide LLC, Irvine, CA) were designed to maintain the soft, natural feel of MemoryGel implants while increasing fullness and projection and minimizing wrinkling, rippling, and related complications through optimization of shell gel-fill.
[OBJECTIVES] To measure 3-year safety and effectiveness of MemoryGel Xtra breast implants in the Mentor MemoryGel and MemoryShape Combined Cohort Clinical Study.
[METHODS] Participants were implanted with MemoryGel Xtra breast implants in a prospective, multicenter clinical trial. Rates of complications and reoperations were analyzed to assess device safety and BREAST-Q was employed to assess device effectiveness.
[RESULTS] Two hundred eighty-seven females receiving MemoryGel Xtra breast implants were enrolled. Complication rates in the primary augmentation cohort included rates of 1.5% for implant-related reoperation, 2.3% for explantation, and 1.5% for Baker grade III or IV capsular contracture. For the revisional augmentation cohort, these rates were 2.8% for implant-related reoperation, 4.3% for explantation, and 3.0% for capsular contracture. For the primary reconstruction cohort, these rates were 12.0% for implant-related reoperation, 12.3% for explantation, and 7.3% for capsular contracture. For the revisional reconstruction cohort, these rates were 7.1% for capsular contracture, with zero implant-related reoperations or explantations. There were no reports of infection or implant malposition or displacement in any of these cohorts. Each cohort showed significantly improved satisfaction with breasts and psychosocial and sexual well-being at 1 year following the primary procedure.
[CONCLUSIONS] These data are consistent with legacy clinical data for MemoryGel and provide the first published safety and effectiveness data regarding the use of MemoryGel Xtra breast implants for breast augmentation and reconstruction.
[OBJECTIVES] To measure 3-year safety and effectiveness of MemoryGel Xtra breast implants in the Mentor MemoryGel and MemoryShape Combined Cohort Clinical Study.
[METHODS] Participants were implanted with MemoryGel Xtra breast implants in a prospective, multicenter clinical trial. Rates of complications and reoperations were analyzed to assess device safety and BREAST-Q was employed to assess device effectiveness.
[RESULTS] Two hundred eighty-seven females receiving MemoryGel Xtra breast implants were enrolled. Complication rates in the primary augmentation cohort included rates of 1.5% for implant-related reoperation, 2.3% for explantation, and 1.5% for Baker grade III or IV capsular contracture. For the revisional augmentation cohort, these rates were 2.8% for implant-related reoperation, 4.3% for explantation, and 3.0% for capsular contracture. For the primary reconstruction cohort, these rates were 12.0% for implant-related reoperation, 12.3% for explantation, and 7.3% for capsular contracture. For the revisional reconstruction cohort, these rates were 7.1% for capsular contracture, with zero implant-related reoperations or explantations. There were no reports of infection or implant malposition or displacement in any of these cohorts. Each cohort showed significantly improved satisfaction with breasts and psychosocial and sexual well-being at 1 year following the primary procedure.
[CONCLUSIONS] These data are consistent with legacy clinical data for MemoryGel and provide the first published safety and effectiveness data regarding the use of MemoryGel Xtra breast implants for breast augmentation and reconstruction.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 해부 | breast
|
유방 | dict | 8 | |
| 합병증 | capsular contracture
|
피막구축 | dict | 4 | |
| 시술 | breast augmentation
|
유방성형술 | dict | 1 | |
| 해부 | capsular
|
scispacy | 1 | ||
| 합병증 | implant-related
|
scispacy | 1 | ||
| 합병증 | infection
|
감염 | dict | 1 | |
| 약물 | [BACKGROUND] Mentor MemoryGel Xtra breast
|
scispacy | 1 | ||
| 약물 | [OBJECTIVES]
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS]
|
scispacy | 1 | ||
| 질환 | fullness
|
C0439650
Fullness
|
scispacy | 1 | |
| 질환 | implant malposition or displacement
|
scispacy | 1 | ||
| 질환 | breasts
|
C0006141
Breast
|
scispacy | 1 | |
| 질환 | LLC
|
scispacy | 1 | ||
| 질환 | MemoryGel Xtra breast
|
scispacy | 1 | ||
| 질환 | Baker grade
|
scispacy | 1 | ||
| 질환 | MemoryGel
|
scispacy | 1 | ||
| 기타 | Irvine
|
scispacy | 1 | ||
| 기타 | MemoryGel implants while increasing fullness and
|
scispacy | 1 | ||
| 기타 | MemoryGel Xtra
|
scispacy | 1 | ||
| 기타 | MemoryShape
|
scispacy | 1 |
MeSH Terms
Female; Humans; Breast Implantation; Breast Implants; Contracture; Follow-Up Studies; Postoperative Complications; Prospective Studies; Prosthesis Design; Reoperation; Silicone Gels
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- The impact of three-dimensional simulation and virtual reality technologies on surgical decision-making and postoperative satisfaction in aesthetic surgery: a preliminary study.
- Cutaneous fistula of the breast: A complication of cosmetic autologous fat transfer.
- Epidermal inclusion cyst after breast reduction mammoplasty.
- Penetrating globe injury following periocular hyaluronic acid filler injection: A case report.
- Clinical outcomes of synthetic absorbable mesh use in breast surgery: First case series in reconstruction and aesthetic mastopexy.