Complications of cosmetic iris implants: French series of 87 eyes.

Journal of cataract and refractive surgery 2020 Vol.46(1) p. 34-39

El Chehab H, Gatinel D, Baudouin C, Muraine M, Hoffart L, Rozot P, Mehanna C, Bonnet C, Nordmann JP, Santiago PY, Burillon C, Baillif S, Pisella PJ, Weber M, Robinet-Perrin A, Deidier D, Hay A, Villain M, Baïkoff G, Gauthier AS, Mathis T, Dot C

Abstract

[PURPOSE] Iris intraocular implants were developed to manage congenital or traumatic iris defects. However, they are also used to change the color of patient eyes. The aim of this retrospective series was to report complications in patients managed in France after cosmetic implantation.

[SETTING] Ophthalmological institutions and private ophthalmologists in France.

[DESIGN] Multicenter retrospective observational study.

[METHODS] Questionnaires were sent to all ophthalmology departments in university hospitals and to private ophthalmologists. This questionnaire listed demographic and clinical data for each implanted eye with a focus on safety, the description of ocular complications (corneal edema, endothelial cell loss, increased intraocular pressure, and intraocular inflammation), and the therapeutic management implemented.

[RESULTS] Forty-four questionnaires (87 eyes) were collected, and ultimately, 33 questionnaires (65 eyes) were considered complete and analyzed. Two types of implants were identified. Of the 65 eyes analyzed, only 5 eyes (7.7%) did not experience any complication and 60 eyes (92.3%) had at least 1 complication. The most commonly reported complication was corneal decompensation (78.5%). The diagnosis of glaucoma was made in over half (52.3%) of the cases. Explantation was needed in 81.5% of cases. The mean final visual acuity was 0.45 ± 0.08 logarithm of the minimum angle of resolution (logMAR) (0 to 2 logMAR).

[CONCLUSIONS] Several ocular complications with a decreased mean visual acuity were described in a young healthy population. In addition, patient information on the safety of this procedure appeared insufficient.

추출된 의학 개체 (NER)

유형영어 표현한국어 / 풀이UMLS CUI출처등장
해부 eye scispacy 1
해부 endothelial cell scispacy 1
해부 intraocular scispacy 1
해부 eyes scispacy 1
해부 corneal scispacy 1
합병증 edema scispacy 1
약물 [PURPOSE] Iris intraocular implants were developed scispacy 1
약물 [DESIGN] Multicenter scispacy 1
약물 [CONCLUSIONS] scispacy 1
질환 traumatic iris defects scispacy 1
질환 ocular complications scispacy 1
질환 corneal edema C0010037
Corneal edema
scispacy 1
질환 endothelial cell loss C0225336
Endothelial Cells
scispacy 1
질환 corneal decompensation C0948284
Corneal decompensation
scispacy 1
질환 glaucoma C0017601
Glaucoma
scispacy 1
기타 iris scispacy 1
기타 patient eyes scispacy 1
기타 patients scispacy 1
기타 patient scispacy 1

MeSH Terms

Adult; Cataract; Corneal Diseases; Female; France; Humans; Iris; Male; Middle Aged; Ocular Hypertension; Postoperative Complications; Prostheses and Implants; Retrospective Studies; Surgery, Plastic; Surveys and Questionnaires; Uveitis, Anterior; Young Adult