Risk Stratification of Thyroid Nodules Using Ultrasound Cine-Loop Video Sequences.
1/5 보강
PICO 자동 추출 (휴리스틱, conf 2/4)
유사 논문P · Population 대상 환자/모집단
189 patients with 329 TNs (78% females, aged 54 ± 15 years; 76% diagnosed via surgery; 14% malignant) were included.
I · Intervention 중재 / 시술
추출되지 않음
C · Comparison 대조 / 비교
추출되지 않음
O · Outcome 결과 / 결론
The integration of cine loops into thyroid ultrasound was found to be a seamless process, thereby enhancing the risk stratification of nodules. Image quality impairments manifested more frequently in cine loops, while static image captures demonstrated higher levels of assessment confidence.
[BACKGROUND/OBJECTIVES] Static image captures (SICs) are the prevailing methodology for documenting thyroid nodules (TNs) on ultrasound examinations.
APA
Schmidt TN, Freesmeyer M, et al. (2025). Risk Stratification of Thyroid Nodules Using Ultrasound Cine-Loop Video Sequences.. Cancers, 17(16). https://doi.org/10.3390/cancers17162616
MLA
Schmidt TN, et al.. "Risk Stratification of Thyroid Nodules Using Ultrasound Cine-Loop Video Sequences.." Cancers, vol. 17, no. 16, 2025.
PMID
40867244 ↗
Abstract 한글 요약
[BACKGROUND/OBJECTIVES] Static image captures (SICs) are the prevailing methodology for documenting thyroid nodules (TNs) on ultrasound examinations. Ultrasound cine-loop (CL) video sequences of the thyroid enable the storage and review of the entire organ in PACS, analogous to sectional imaging modalities such as CT or MRI. Expanding on SIC, the collection of more extensive datasets is possible, with the potential to enhance diagnostic performance. However, there is an absence of reliable data concerning this process.
[METHODS] This retrospective, tertiary care, single-center study included all patients with cytologically and histopathologically diagnosed TNs from 01/16 to 12/23. A standardized CL protocol was routinely acquired in addition to conventional SIC. The diagnostic performance of ACR and Kwak TIRADS was examined for both CL and SIC in a PACS. Advantages, challenges, and limitations of CL were analyzed.
[CONCLUSIONS] In total, 189 patients with 329 TNs (78% females, aged 54 ± 15 years; 76% diagnosed via surgery; 14% malignant) were included. On SIC, 58 TNs (18%) were not identified (all benign). When comparing CL with SIC, a strong correlation was observed for nearly all ultrasound features (echogenicity, composition, margin, and foci; each < 0.001) and both TIRADSs (each < 0.001). The diagnostic accuracy of CL was slightly superior, with maximum values of 85% for ACR and 87% for Kwak TIRADS, respectively. Rating confidence and image quality exhibited superiority on SIC (each < 0.001). The occurrence of image artifacts was more prevalent in CL ( < 0.001). The integration of cine loops into thyroid ultrasound was found to be a seamless process, thereby enhancing the risk stratification of nodules. Image quality impairments manifested more frequently in cine loops, while static image captures demonstrated higher levels of assessment confidence.
[METHODS] This retrospective, tertiary care, single-center study included all patients with cytologically and histopathologically diagnosed TNs from 01/16 to 12/23. A standardized CL protocol was routinely acquired in addition to conventional SIC. The diagnostic performance of ACR and Kwak TIRADS was examined for both CL and SIC in a PACS. Advantages, challenges, and limitations of CL were analyzed.
[CONCLUSIONS] In total, 189 patients with 329 TNs (78% females, aged 54 ± 15 years; 76% diagnosed via surgery; 14% malignant) were included. On SIC, 58 TNs (18%) were not identified (all benign). When comparing CL with SIC, a strong correlation was observed for nearly all ultrasound features (echogenicity, composition, margin, and foci; each < 0.001) and both TIRADSs (each < 0.001). The diagnostic accuracy of CL was slightly superior, with maximum values of 85% for ACR and 87% for Kwak TIRADS, respectively. Rating confidence and image quality exhibited superiority on SIC (each < 0.001). The occurrence of image artifacts was more prevalent in CL ( < 0.001). The integration of cine loops into thyroid ultrasound was found to be a seamless process, thereby enhancing the risk stratification of nodules. Image quality impairments manifested more frequently in cine loops, while static image captures demonstrated higher levels of assessment confidence.
🏷️ 키워드 / MeSH 📖 같은 키워드 OA만
🏷️ 같은 키워드 · 무료전문 — 이 논문 MeSH/keyword 기반
- Clinical Presentation and Outcomes of Patients Undergoing Surgery for Thyroid Cancer.
- Microfluidic-based patient-derived organoids recapitulate thyroid cancer heterogeneity and reveal NF-κB-driven maturation for precision therapy.
- Association of patient health education with the postoperative health related quality of life in low- intermediate recurrence risk differentiated thyroid cancer patients.
- CHI3L1 Is Associated With TP53 Signaling and Promotes Papillary Thyroid Carcinoma Progression.
- DNA Damage and Repair in Thyroid Physiology and Disease.
- Impact of health information prescription in thyroid cancer.