Evaluation of the safety and efficacy of the Sentire Surgical System (C1000) for robot-assisted radical prostatectomy.
1/5 보강
PICO 자동 추출 (휴리스틱, conf 2/4)
유사 논문P · Population 대상 환자/모집단
20 patients were recruited.
I · Intervention 중재 / 시술
추출되지 않음
C · Comparison 대조 / 비교
추출되지 않음
O · Outcome 결과 / 결론
Seventeen patients achieved undetectable levels of prostate-specific antigen at 1 month after surgery. [CONCLUSION] These initial results support the Sentire Surgical System (C1000), Hong Kong's first locally developed multidisciplinary robotic surgical platform, as a safe and effective option for radical prostatectomy, with clinical performance and outcomes comparable to existing robotic systems.
[INTRODUCTION] This prospective clinical study evaluated the clinical safety and efficacy of the Sentire Surgical System (C1000), a locally developed robotic surgical platform, in performing radical p
APA
Ng CF, Yee CH, et al. (2025). Evaluation of the safety and efficacy of the Sentire Surgical System (C1000) for robot-assisted radical prostatectomy.. Hong Kong medical journal = Xianggang yi xue za zhi, 31(3), 192-198. https://doi.org/10.12809/hkmj2412169
MLA
Ng CF, et al.. "Evaluation of the safety and efficacy of the Sentire Surgical System (C1000) for robot-assisted radical prostatectomy.." Hong Kong medical journal = Xianggang yi xue za zhi, vol. 31, no. 3, 2025, pp. 192-198.
PMID
40461168 ↗
Abstract 한글 요약
[INTRODUCTION] This prospective clinical study evaluated the clinical safety and efficacy of the Sentire Surgical System (C1000), a locally developed robotic surgical platform, in performing radical prostatectomy in Hong Kong.
[METHODS] This was a single-centre, single-arm study. Adult patients with a clinical diagnosis of localised prostate cancer planned for surgical treatment were invited to participate. Surgery was performed using the Sentire C1000 system following the standard approach. The primary endpoints were the conversion rate and the incidence of perioperative complications within 30 days. Secondary outcomes, including perioperative, pathological, oncological, and functional outcomes at 1 month after surgery, were also assessed.
[RESULTS] From August 2022 to September 2023, 20 patients were recruited. All procedures were performed without conversion. There were no intraoperative complications related to the robotic device. Minor surgical complications (Grade I-II according to the Clavien-Dindo Classification) occurred in seven patients and were managed conservatively. The mean total operative time was 184.5 minutes (standard deviation=30.0). The median estimated blood loss was 175.0 mL (interquartile range [IQR]=100.0-275.0). The median length of hospital stay was 3.0 days (IQR=2.0-4.0). Seventeen patients achieved undetectable levels of prostate-specific antigen at 1 month after surgery.
[CONCLUSION] These initial results support the Sentire Surgical System (C1000), Hong Kong's first locally developed multidisciplinary robotic surgical platform, as a safe and effective option for radical prostatectomy, with clinical performance and outcomes comparable to existing robotic systems.
[METHODS] This was a single-centre, single-arm study. Adult patients with a clinical diagnosis of localised prostate cancer planned for surgical treatment were invited to participate. Surgery was performed using the Sentire C1000 system following the standard approach. The primary endpoints were the conversion rate and the incidence of perioperative complications within 30 days. Secondary outcomes, including perioperative, pathological, oncological, and functional outcomes at 1 month after surgery, were also assessed.
[RESULTS] From August 2022 to September 2023, 20 patients were recruited. All procedures were performed without conversion. There were no intraoperative complications related to the robotic device. Minor surgical complications (Grade I-II according to the Clavien-Dindo Classification) occurred in seven patients and were managed conservatively. The mean total operative time was 184.5 minutes (standard deviation=30.0). The median estimated blood loss was 175.0 mL (interquartile range [IQR]=100.0-275.0). The median length of hospital stay was 3.0 days (IQR=2.0-4.0). Seventeen patients achieved undetectable levels of prostate-specific antigen at 1 month after surgery.
[CONCLUSION] These initial results support the Sentire Surgical System (C1000), Hong Kong's first locally developed multidisciplinary robotic surgical platform, as a safe and effective option for radical prostatectomy, with clinical performance and outcomes comparable to existing robotic systems.
🏷️ 키워드 / MeSH 📖 같은 키워드 OA만
🏷️ 같은 키워드 · 무료전문 — 이 논문 MeSH/keyword 기반
- A Phase I Study of Hydroxychloroquine and Suba-Itraconazole in Men with Biochemical Relapse of Prostate Cancer (HITMAN-PC): Dose Escalation Results.
- Self-management of male urinary symptoms: qualitative findings from a primary care trial.
- Clinical and Liquid Biomarkers of 20-Year Prostate Cancer Risk in Men Aged 45 to 70 Years.
- Diagnostic accuracy of Ga-PSMA PET/CT versus multiparametric MRI for preoperative pelvic invasion in the patients with prostate cancer.
- Clinical Presentation and Outcomes of Patients Undergoing Surgery for Thyroid Cancer.
- Association of patient health education with the postoperative health related quality of life in low- intermediate recurrence risk differentiated thyroid cancer patients.