An automated method to establish patient specific asymmetric intrafraction motion monitoring tolerances for prostate radiotherapy.
1/5 보강
[PURPOSE] Intrafraction motion management is recognized as a critical component of radiotherapy.
APA
Beeksma B, Daniel J, et al. (2026). An automated method to establish patient specific asymmetric intrafraction motion monitoring tolerances for prostate radiotherapy.. Journal of applied clinical medical physics, 27(3), e70507. https://doi.org/10.1002/acm2.70507
MLA
Beeksma B, et al.. "An automated method to establish patient specific asymmetric intrafraction motion monitoring tolerances for prostate radiotherapy.." Journal of applied clinical medical physics, vol. 27, no. 3, 2026, pp. e70507.
PMID
41755679 ↗
Abstract 한글 요약
[PURPOSE] Intrafraction motion management is recognized as a critical component of radiotherapy. While clinical trials often recommend its use, they rarely define motion management tolerances. Consequently, institutions set tolerance levels independently, often without considering patient specific anatomy, dose distributions or direction of motion. This study introduces an automated method that can be used to derive patient specific, asymmetric intrafraction motion monitoring tolerances based on individualized treatment plans.
[METHODS] Treatment plans for 20 prostate cancer patients receiving a simultaneous integrated boost to dominant intraprostatic lesions (DIL) were retrospectively analyzed. Referred to as the isocentre translation method, patient movement was simulated by recalculating plans with the isocentre shifted by 2, 4 and 5 mm in six different directions. The magnitude and direction that resulted in violations of dose constraints for organ at risk (OAR) rectum, bladder and urethra was determined assuming a systematic shift for all fractions. A second automated approach, the contour translation method, used Eclipse Scripting API to estimate directional tolerances via contour translation and Boolean operations, avoiding dose recalculation. Results for this method were validated against the isocentre translation method and compared for efficiency by assessing processing time.
[RESULTS] The magnitude and direction of motion to cause OAR constraint violations were organ and patient specific. Urethral violations were most sensitive to shifts toward the DIL, whereas bladder and rectum constraints were primarily affected by anterior/superior and posterior shifts, respectively. The contour translation method was demonstrated to be equivalent to the isocentre translation method within an equivalence margin of 0.5 mm. The contour translation method significantly reduced processing time (4 min 40s vs. 47 min 52s per patient).
[CONCLUSION] The direction and extent of motion impacting OAR constraints vary by patient and organ, supporting the need for personalized intrafraction motion monitoring tolerances. The proposed contour translation method provides a practical, efficient process that is able to facilitate individualized motion management in clinical workflows.
[METHODS] Treatment plans for 20 prostate cancer patients receiving a simultaneous integrated boost to dominant intraprostatic lesions (DIL) were retrospectively analyzed. Referred to as the isocentre translation method, patient movement was simulated by recalculating plans with the isocentre shifted by 2, 4 and 5 mm in six different directions. The magnitude and direction that resulted in violations of dose constraints for organ at risk (OAR) rectum, bladder and urethra was determined assuming a systematic shift for all fractions. A second automated approach, the contour translation method, used Eclipse Scripting API to estimate directional tolerances via contour translation and Boolean operations, avoiding dose recalculation. Results for this method were validated against the isocentre translation method and compared for efficiency by assessing processing time.
[RESULTS] The magnitude and direction of motion to cause OAR constraint violations were organ and patient specific. Urethral violations were most sensitive to shifts toward the DIL, whereas bladder and rectum constraints were primarily affected by anterior/superior and posterior shifts, respectively. The contour translation method was demonstrated to be equivalent to the isocentre translation method within an equivalence margin of 0.5 mm. The contour translation method significantly reduced processing time (4 min 40s vs. 47 min 52s per patient).
[CONCLUSION] The direction and extent of motion impacting OAR constraints vary by patient and organ, supporting the need for personalized intrafraction motion monitoring tolerances. The proposed contour translation method provides a practical, efficient process that is able to facilitate individualized motion management in clinical workflows.
🏷️ 키워드 / MeSH 📖 같은 키워드 OA만
- Humans
- Prostatic Neoplasms
- Male
- Radiotherapy Planning
- Computer-Assisted
- Radiotherapy Dosage
- Radiotherapy
- Intensity-Modulated
- Organs at Risk
- Retrospective Studies
- Movement
- Algorithms
- Automation
- asymmetric motion management
- dominant intraprostatic lesion
- image‐guided radiotherap
- intrafraction motion
- patient specific tolerances
- prostate radiotherapy
- treatment planning automation
🏷️ 같은 키워드 · 무료전문 — 이 논문 MeSH/keyword 기반
- A Phase I Study of Hydroxychloroquine and Suba-Itraconazole in Men with Biochemical Relapse of Prostate Cancer (HITMAN-PC): Dose Escalation Results.
- Self-management of male urinary symptoms: qualitative findings from a primary care trial.
- Clinical and Liquid Biomarkers of 20-Year Prostate Cancer Risk in Men Aged 45 to 70 Years.
- Diagnostic accuracy of Ga-PSMA PET/CT versus multiparametric MRI for preoperative pelvic invasion in the patients with prostate cancer.
- Association of patient health education with the postoperative health related quality of life in low- intermediate recurrence risk differentiated thyroid cancer patients.
- Early local immune activation following intra-operative radiotherapy in human breast tissue.